Categories
Nevin Manimala Statistics

Is There a Difference in Screw Accuracy, Robot Time Per Screw, Robot Abandonment, and Radiation Exposure Between the Mazor X and the Renaissance? A Propensity-Matched Analysis of 1179 Robot-Assisted Screws

Global Spine J. 2021 Jul 8:21925682211029867. doi: 10.1177/21925682211029867. Online ahead of print.

ABSTRACT

STUDY DESIGN: Prospective single-cohort analysis.

OBJECTIVES: To compare the outcomes/complications of 2 robotic systems for spine surgery.

METHODS: Adult patients (≥18-years-old) who underwent robot-assisted spine surgery from 2016-2019 were assessed. A propensity score matching (PSM) algorithm was used to match Mazor X to Renaissance cases. Preoperative CT scan for planning and an intraoperative O-arm for screw evaluation were preformed. Outcomes included screw accuracy, robot time/screw, robot abandonment, and radiation. Screw accuracy was measured using Vitrea Core software by 2 orthopedic surgeons. Screw breach was measured according to the Gertzbein/Robbins classification.

RESULTS: After PSA, a total of 65 patients (Renaissance: 22 vs. X: 43) were included. Patient/operative factors were similar between robot systems (P > .05). The pedicle screw accuracy was similar between robots (Renaissance: 1.1%% vs. X: 1.3%, P = .786); however, the S2AI screw breach rate was significantly lower for the X (Renaissance: 9.5% vs. X: 1.2%, P = .025). Robot time per screw was not statistically different (Renaissance: 4.6 minutes vs. X: 3.9 minutes, P = .246). The X was more reliable with an abandonment rate of 2.3% vs. Renaissance:22.7%, P = .007. Radiation exposure were not different between robot systems. Non-robot related complications including dural tear, loss of motor/sensory function, and blood transfusion were similar between robot systems.

CONCLUSION: This is the first comparative analyses of screw accuracy, robot time/screw, robot abandonment, and radiation exposure between the Mazor X and Renaissance systems. There are substantial improvements in the X robot, particularly in the perioperative planning processes, which likely contribute to the X’s superiority in S2AI screw accuracy by nearly 8-fold and robot reliability by nearly 10-fold.

PMID:34235996 | DOI:10.1177/21925682211029867