J Chromatogr A. 2021 Sep 22;1657:462570. doi: 10.1016/j.chroma.2021.462570. Online ahead of print.
ABSTRACT
The purpose of this paper is to propose a simple method for daily inspections of a gas chromatography-mass spectrometry (GC-MS) system with an instrumental detection limit (IDL) as an indicator. A definition of DLs by ISO is 3.3σ where σ denotes the standard deviation (SD) of blank measurements. Estimation of σ is carried out according to the function of mutual information (FUMI) theory and actually with commercial software (TOCO19). An IDL which is a combination of a signal area, width and noise level is concluded to be a good indicator for daily inspections compared with each of its constituents. Methyl stearate is used as a standard material for the daily inspection of a programmed-temperature GC-MS system. A short chromatogram of 1800 data points (1.5 min) containing a target signal and background noise is fit for the IDL prediction by TOCO19. The relative SDs (RSDs) based on the theoretically estimated σ are shown to coincide with statistical results from repeated measurements within 95% confidence intervals. Column temperature is observed to affect IDLs through background fluctuation and then temperature-IDL relationship is examined in a range from 170 to 270 °C. Actual daily inspections over a month are demonstrated.
PMID:34607291 | DOI:10.1016/j.chroma.2021.462570