Radioablation of Upper Abdominal Malignancies by CT-Guided, Interstitial HDR Brachytherapy: A Multivariate Analysis of Catheter Placement Assisted by Ultrasound Imaging

Rofo. 2021 Oct 14. doi: 10.1055/a-1545-4983. Online ahead of print.

ABSTRACT

PURPOSE: To evaluate the use of ultrasound (US) during catheter placement in interstitial brachytherapy (iBT) of abdominal malignancies as an alternative to computed tomography (CT) fluoroscopy.

MATERIALS AND METHODS: Catheter placement for CT-guided iBT was, if US visibility was sufficient, assisted by sonography in 52 consecutive patients with 82 lesions (liver N = 62; kidney N = 16; peritoneum N = 4) of various malignancies. We collected data on lesion visibility, location, depth, size, and dosimetry. Comparison of CT fluoroscopy versus US-assisted catheter placement was performed by Fisher’s exact test for frequencies and U-test for lesion visibility and dosimetric data. Factors predicting the utility of sonography were determined in a lesion-based multivariate regression analysis. A p-value < 0.05 was regarded as statistically significant.

RESULTS: 150 catheters (1 to 6 per lesion; mean diameter 3.6 ± 2.4 cm) were implanted. CT fluoroscopy was used for 44 catheters, and US was used for 106 catheters. Lesion visibility assessed by 5-point Likert scale was significantly better in US (median 2 vs. 3; p = 0.011) and effective dose was significantly reduced if US guidance was applicable (median 1.75 vs. 8.19 mSv; p = 0.014). In a multivariate regression analysis, we identified increased lesion size and caudal location within the target organ to independently predict the utility of ultrasound in catheter placement for iBT.

CONCLUSION: Sonography is a helpful technique to assist CT-guided interstitial brachytherapy of upper abdominal malignancies. Especially for larger lesions localized in the lower liver segments or lower half of the kidney, superior visibility can be expected. As the effective dose of the patient is also reduced, radiation exposure of the medical staff may be indirectly lowered.

KEY POINTS: · Ultrasound-assisted catheter placement in CT-guided brachytherapy of upper abdominal malignancies significantly improves lesion visibility.. · Predictors of successful ultrasound application are larger lesions within the lower portion of the liver and kidney.. · By reducing the need for CT fluoroscopy during intervention, radiation exposure to the medical staff may be indirectly lowered..

CITATION FORMAT: · Damm R, Damm R, Heinze C et al. Radioablation of Upper Abdominal Malignancies by CT-Guided, Interstitial HDR Brachytherapy: A Multivariate Analysis of Catheter Placement Assisted by Ultrasound Imaging. Fortschr Röntgenstr 2021; DOI: 10.1055/a-1545-4983.