Categories
Nevin Manimala Statistics

Feasibility of the automated column agglutination technique for titration of anti-A/B antibodies in ABO-incompatible living kidney transplantation

Ther Apher Dial. 2021 Nov 23. doi: 10.1111/1744-9987.13764. Online ahead of print.

ABSTRACT

INTRODUCTION: Quantitative measurement of anti-A/-B antibody titers is important during ABO-incompatible living kidney transplantation (ABOi-LKT).

METHODS: We conducted a multi-institutional study to measure the antibody titers using the automated column agglutination technique (auto-CAT) and tube test (TT) method in ABOi-LKT recipients. Statistical analysis was performed to evaluate the two methods.

RESULTS: We examined 111 samples from 35 ABOi-LKT recipients at four institutions. The correlation coefficient of the two methods was >0.9; the concordance rate and clinically acceptable concordance rate for the IgG titers were 60.4% and 88.3%, respectively. Perioperative status did not influence the statistical significance. Parallel changes were observed in the IgG antibody titers measured using the auto-CAT or TT technique by desensitizing therapy in time-course monitoring.

CONCLUSION: Auto-CAT is comparable with the TT technique and is feasible for IgG anti-A/B antibody titration in ABOi-LKT recipients. This article is protected by copyright. All rights reserved.

PMID:34812590 | DOI:10.1111/1744-9987.13764

By Nevin Manimala

Portfolio Website for Nevin Manimala