Synthesis of pseudo-CT images from pelvic MRI images based on MD-CycleGAN model for radiotherapy

Phys Med Biol. 2021 Dec 8. doi: 10.1088/1361-6560/ac4123. Online ahead of print.

ABSTRACT

OBJECTIVE: A multi-discriminator-based cycle generative adversarial network (MD-CycleGAN) model was proposed to synthesize higher-quality pseudo-CT from MRI.

APPROACH: The MRI and CT images obtained at the simulation stage with cervical cancer were selected to train the model. The generator adopted the DenseNet as the main architecture. The local and global discriminators based on convolutional neural network jointly discriminated the authenticity of the input image data. In the testing phase, the model was verified by four-fold cross-validation method. In the prediction stage, the data were selected to evaluate the accuracy of the pseudo-CT in anatomy and dosimetry, and they were compared with the pseudo-CT synthesized by GAN with generator based on the architectures of ResNet, sU-Net, and FCN.

MAIN RESULTS: There are significant differences(P<0.05) in the four-fold-cross validation results on peak signal-to-noise ratio and structural similarity index metrics between the pseudo-CT obtained based on MD-CycleGAN and the ground truth CT (CTgt). The pseudo-CT synthesized by MD-CycleGAN had closer anatomical information to the CTgt with root mean square error of 47.83±2.92 HU and normalized mutual information value of 0.9014±0.0212 and mean absolute error value of 46.79±2.76 HU. The differences in dose distribution between the pseudo-CT obtained by MD-CycleGAN and the CTgt were minimal. The mean absolute dose errors of Dosemax, Dosemin and Dosemean based on the planning target volume were used to evaluate the dose uncertainty of the four pseudo-CT. The u-values of the Wilcoxon test were 55.407, 41.82 and 56.208, and the differences were statistically significant. The 2%/2 mm-based gamma pass rate (%) of the proposed method was 95.45±1.91, and the comparison methods (ResNet_GAN, sUnet_GAN and FCN_GAN) were 93.33±1.20, 89.64±1.63 and 87.31±1.94, respectively.

SIGNIFICANCE: The pseudo-CT obtained based on MD-CycleGAN have higher imaging quality and are closer to the CTgt in terms of anatomy and dosimetry than other GAN models.

PMID:34879356 | DOI:10.1088/1361-6560/ac4123