Quantitative evaluation of prospective motion correction in healthy subjects at 7T MRI

Magn Reson Med. 2021 Aug 31. doi: 10.1002/mrm.28998. Online ahead of print.

ABSTRACT

PURPOSE: Quantitative assessment of prospective motion correction (PMC) capability at 7T MRI for compliant healthy subjects to improve high-resolution images in the absence of intentional motion.

METHODS: Twenty-one healthy subjects were imaged at 7 T. They were asked not to move, to consider only unintentional motion. An in-bore optical tracking system was used to monitor head motion and consequently update the imaging volume. For all subjects, high-resolution T1 (3D-MPRAGE), T2 (2D turbo spin echo), proton density (2D turbo spin echo), and T2∗ (2D gradient echo) weighted images were acquired with and without PMC. The images were evaluated through subjective and objective analysis.

RESULTS: Subjective evaluation overall has shown a statistically significant improvement (5.5%) in terms of image quality with PMC ON. In a separate evaluation of every contrast, three of the four contrasts (T1 , T2 , and proton density) have shown a statistically significant improvement (9.62%, 9.85%, and 9.26%), whereas the fourth one ( T2∗ ) has shown improvement, although not statistically significant. In the evaluation with objective metrics, average edge strength has shown an overall improvement of 6% with PMC ON, which was statistically significant; and gradient entropy has shown an overall improvement of 2%, which did not reach statistical significance.

CONCLUSION: Based on subjective assessment, PMC improved image quality in high-resolution images of healthy compliant subjects in the absence of intentional motion for all contrasts except T2∗ , in which no significant differences were observed. Quantitative metrics showed an overall trend for an improvement with PMC, but not all differences were significant.

PMID:34463376 | DOI:10.1002/mrm.28998