Biomed Phys Eng Express. 2022 Jan 17. doi: 10.1088/2057-1976/ac4c2a. Online ahead of print.
ABSTRACT
By using the statistical techniques of ANOVA: mean test and regression, it was found that the NKRcalibration factor of Standard Imaging (SI) model HDR 1000 plus chambers presents a quadratic dependence with the reference air kerma rateKR(from 6.9 mGy h-1to 43.9 mGy h-1). In order to understand and correct this dependency one model is presented for total recombination:ks=I300/I150=1+kini+ kd+kvolI300+kscreenI3002, wherekiniis the initial recombination,kdthe thermal diffusion recombination,kvolthe volumetric recombination andkscreenthe screening for the currents/charges collected at the potential differences of 300 and 150 V. In conclusion, the total recombinationksis composed by onekiniwith a constant contribution of 0.019%, onekdcontribution of 0.017%, onekvolI300contribution from 0.022% to 0.138%, and thekscreenI3002effects from 0.002% to 0.09% in the range ofKRrate above. However, when this model forksis applied to try to correct the quadratic dependence of theNKRvs KR, explicitly there is no improvement in the variation range of 0.5% of theNKRvs KR.Nonetheless, it allows to obtainNKRvalues consistent with a uc≤ 0.7 %, which is less than 1.25% reported in the literature by ADCL or SSDL.
PMID:35039465 | DOI:10.1088/2057-1976/ac4c2a