Phys Eng Sci Med. 2022 Oct 13. doi: 10.1007/s13246-022-01184-6. Online ahead of print.
ABSTRACT
This work details the clinical pilot study methodology used at Wellington Blood and Cancer Centre (WBCC) before the clinical release of in vivo dosimetry (IVD) system EPIgray™ for head and neck (H&N) volumetric modulated arc therapy (VMAT) treatments. Clinical pilot studies make it possible to select appropriate, department-specific tolerance ranges for the treatment type and site under investigation. An IVD clinical pilot study of H&N VMAT treatments was conducted over 3 months at WBCC using EPIgray™ dose reconstruction software and included 12 patients and 32 individual treatment fractions. Statistical analysis of the dose deviations between the treatment planning system (TPS) dose and EPIgray™ reconstructed dose confirmed that a deviation tolerance range of ± 7.0% was an appropriate choice for H&N VMAT at WBCC.
PMID:36227496 | DOI:10.1007/s13246-022-01184-6