Orthop Res Rev. 2025 May 7;17:189-198. doi: 10.2147/ORR.S514237. eCollection 2025.
ABSTRACT
STUDY DESIGN: Retrospective chart review.
OBJECTIVE: This study aims to investigate the application of robots in treating adult degenerative scoliosis (ADS) with circumferential minimal invasive surgery (cMIS).
BACKGROUND: The cMIS is useful but faces a lot of challenges in correcting ADS. One of the most important challenges is the difficulty in screw placement. Robot-assisted technique demonstrates lots of advantages but the data about its application in treating ADS is limited in literatures.
METHODS: A total of 51 cases diagnosed with ADS were retrospectively analyzed. All patients underwent cMIS technique with staged surgeries (OLIF and PPS fixation). Group A enrolled 21 patients and performed robot-assisted technique. Group B enrolled 30 patients and performed fluoroscopy guided technique. Clinical outcomes like the operation time, radiation exposure, pressure curve and post-operation VAS score were recorded. 3D-CT scan was also performed to evaluate the accuracy of the screws.
RESULTS: The average preparation time were much higher in group A (23.4 ± 2.8 vs 3.1 ± 1.0 min, p < 0.0001). But the total operation time was similar (62.7 ± 12.5 vs 55.7 ± 20.6 min, p = 0.174). The average fluoroscopic scan number were 9.4 ± 1.7 in group A, much lower than that of group A (27.7 ±5.9, p < 0.001). No statistical difference was found with the VAS scale between the groups (p = 0.631). No matter considers only screws of grade A as perfect screws (81.5% vs 73.8%) or considers both grade A and B as acceptable screws (93.8% vs 87.7%), group A demonstrated significant higher screw accuracy (p = 0.038, p = 0.018, respectively). Also, the robots demonstrated significant less facet joint violence (p < 0.0001), larger inward tilt angle (p < 0.0001), and longer screw length (p = 0.0008).
CONCLUSION: The robot-assisted technique demonstrated significant advantages like higher pedicle screw accuracy, better trajectory, less radiation exposure, but similar operation time compared with fluoroscopy guided technique in treating ADS with CMIS.
PMID:40357444 | PMC:PMC12067674 | DOI:10.2147/ORR.S514237