Feasibility of a Video Otoscope for Diagnosis of Otologic Pathology in the Pediatric Emergency Department

Pediatr Emerg Care. 2023 Jun 12. doi: 10.1097/PEC.0000000000002988. Online ahead of print.

ABSTRACT

OBJECTIVES: Performing pediatric otoscopy can be difficult secondary to patient compliance, which potentiates misdiagnosis and inaccurate treatment of acute otitis media. This study used a convenience sample to assess the feasibility of using a video otoscope for the examination of tympanic membranes in children presenting to a pediatric emergency department.

METHODS: We obtained otoscopic videos using the JEDMED Horus + HD Video Otoscope. Participants were randomized to video or standard otoscopy, and a physician completed their bilateral ear examinations. In the video group, physicians reviewed otoscope videos with the patient’s caregiver. The caregiver and physician completed separate surveys using a 5-point Likert Scale regarding perceptions of the otoscopic examination. A second physician reviewed each otoscopic video.

RESULTS: We enrolled 213 participants in 2 groups (standard otoscopy, n = 94; video otoscopy, n = 119). We used Wilcoxon rank sum, Fisher exact test, and descriptive statistics to compare results across groups. For physicians, there were no statistically significant differences between groups with ease of device use, quality of otoscopic view, or diagnosis. There was moderate agreement between physician video otoscopic view satisfaction and slight agreement between physician video otologic diagnosis. Estimates of length of time to complete the ear examinations were longer more often for the video otoscope compared with standard for both caregivers (OR, 2.00; 95% confidence interval, 1.10-3.70; P = 0.02) and physicians (OR, 3.08; 95% confidence interval, 1.67-5.78; P < 0.01). There were no statistically significant differences between video and standard otoscopy with regard to caregiver perception of comfort, cooperation, satisfaction, or diagnosis understanding.

CONCLUSIONS: Caregivers perceive that video otoscopy and standard otoscopy are comparable in comfort, cooperation, examination satisfaction, and diagnosis understanding. Physicians made a wider range of more subtle diagnoses with the video otoscope. However, examination length of time may limit the JEDMED Horus + HD Video Otoscope’s feasibility in a busy pediatric emergency department.

PMID:37308169 | DOI:10.1097/PEC.0000000000002988