Biomechanical Evaluation of the 2 Different Levels of Coracoid Graft Positions in the Latarjet Procedure for Anterior Shoulder Instability

Orthop J Sports Med. 2023 Dec 22;11(12):23259671231202533. doi: 10.1177/23259671231202533. eCollection 2023 Dec.

ABSTRACT

BACKGROUND: In the Latarjet procedure, the ideal placement of the coracoid graft in the medial-lateral position is flush with the anterior glenoid rim. However, the ideal position of the graft in the superior-inferior position (sagittal plane) for restoring glenohumeral joint stability is still controversial.

PURPOSE: To compare coracoid graft clockface positions between the traditional 3 to 5 o’clock and a more inferior (for the right shoulder) 4 to 6 o’clock with regard to glenohumeral joint stability in the Latarjet procedure.

STUDY DESIGN: Controlled laboratory study.

METHODS: A total of 10 fresh-frozen cadaveric shoulders were tested in a dynamic, custom-built robotic shoulder model. Each shoulder was loaded with a 50-N compressive load while an 80-N force was applied in the anteroinferior axes at 90° of abduction and 60° of shoulder external rotation. Four conditions were tested: (1) intact, (2) 6-mm glenoid bone loss (GBL), (3) Latarjet procedure fixed at 3- to 5-o’clock position, and (4) Latarjet procedure fixed at 4- to 6-o’clock position. The stability ratio (SR) and degree of lateral humeral displacement (LHD) were recorded. A 1-factor random-intercepts linear mixed-effects model and Tukey method were used for statistical analysis.

RESULTS: Compared with the intact state (1.77 ± 0.11), the SR was significantly lower after creating a 6-mm GBL (1.14 ± 0.61, P = .009), with no significant difference in SR after Latarjet 3 to 5 o’clock (1.51 ± 0.70, P = .51) or 4 to 6 o’clock (1.55 ± 0.68, P = .52). Compared with the intact state (6.48 ± 2.24 mm), LHD decreased significantly after GBL (3.16 ± 1.56 mm, P < .001) and Latarjet 4 to 6 o’clock (5.48 ± 3.39 mm, P < .001). Displacement decreased significantly after Latarjet 3 to 5 o’clock (4.78 ± 2.50 mm, P = .04) compared with the intact state but not after Latarjet 4 to 6 o’clock (P = .71).

CONCLUSION: The Latarjet procedure in both coracoid graft positions (3-5 and 4-6 o’clock) restored the SR to the values measured in the intact state. A more inferior graft position (fixed at 4-6 o’clock) may improve shoulder biomechanics, but additional work is needed to establish clinical relevance.

CLINICAL RELEVANCE: An inferior coracoid graft fixation, the 4- to 6-o’clock position, may benefit in restoring normal shoulder biomechanics after the Latarjet procedure.

PMID:38145219 | PMC:PMC10748952 | DOI:10.1177/23259671231202533